10. Calibration Procedure

i
• This procedure should be performed in the same
geographic area as the measurements for infant weights.
Procedure requires two certified 10 kg Class F test weights.
1. Press
button while the scale is off.
2. Press
1
button while the display
is flashing all segments --> Display
shows "Func 1".
reweigh
fed baby
3. Press
button --> Display
shows "Func 2".
4. Press
button --> Display shows
3
"9.XXX" (For example, 9.802 will be
displayed).
i
• The gravity setting will update automatically upon
completing the calibration procedure.