Technical Specifications

Titrator TitroLine
1.3.1
Translation of the legally binding German version
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1: 2013
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: 2010 for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Country of origin: Germany, Made in Germany
The following solvents/titration reagents are allowed to be used:
 All common titration solutions.
 As reagent water and all non-aggressive non-organic and organic fluids are allowed.
 If using combustible fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the
chemical industry.
 For fluids with higher viscosity ( 5 mm
speed can be adjusted.
 Fluids with viscosity over 20mm
To ensure maximum accuracy of the readings we recommend to allow some reasonable time for the
®
5000 to "warm up".
TitroLine
Measuring input 1 (analog):
pH/mV-input with 12 bit transducer for high-precision readings.
Electrode socket according to DIN 19 262 or additional with BNC socket insert (Z 860).
Reference electrode 1 x 4 mm socket.
pH
pH
U mV
mV
Measuring input (Pt 1000):
Temperature sensor - connector for a resistance thermometer Pt 1000 and NTC 30 kOhm.
Connection: 2 x 4 mm sockets.
Display:
3.5 inches -1/4 VGA TFT display with 320x240 pixels.
Calibration:
Automatically with up to three buffer solutions, sequence during calibration optional, freely
definable buffers can be input.
Default buffer solutions according to DIN 19 266 and NBS, or technical buffers:
pH = 1.00; pH = 4.00; pH = 4.01; pH = 6.87; pH= 7.00; pH = 9.18; pH = 10.00
Input:
Measurement input 1: pH/mV-input with electrode socket according DIN 19 262/or BNC
Measurement input Pt 1000: Temperature sensor probe for resistance thermometer Pt 1000
Power supply:
power supply 100-240 V; 50/60 Hz, power input: 30 VA
* The measurement uncertainty of the sensor probe has to be taken into account as well.
®
5000
2
/s), lower boiling point or affinity to outgas, the filling and dosage
2
/s cannot be dosed.
Measurement range Display resolution
- 3.0 ... 17.00
- 1900 ... 1900
Measurement
range T [°C]
Pt 1000
- 30 ... 115
NTC 30
- 30 ... 115
(Connection sockets: 2 x 4 mm)
Use the power supply TZ 1853 only!
Measurement accuracy*
without sensor probe
0.01
0.05
1
1.0
Measurement accuracy* without
Display resolution
0.1
0.1
(Release: 18. June 2020)
Input
resistance 
 1 Digit
> 5  10
 1 Digit
> 5  10
sensor probe
0.5 K  1 Digit
0.5 K  1 Digit
12
12