DPOAE
Stimulus
Frequency range:
Frequency step:
Level:
Level Step:
Transducer:
Recording
Analysis time:
A/D Resolution:
Artifact rejection system:
Stimulus tolerance:
SNR criteria:
Probe check window:
DP-response window:
Display
General display gain:
Display
Probe specifications
Titan IOW probe:
Other
Test Pressure:
General
PC control:
Memory:
Thermal printer
(Optional):
Dimensions
Titan Weight
ABR/OAE/IMP
shoulderbox weight
ABR/OAE/IMP
shoulderbox
Dimensions
OAE/IMP shoulderbox
weight
OAE/IMPshoulderbox
Dimensions
Titan Instruction for Use - English
Date: 2011-06-06
USB:
Type:
Paper width:
Printing time:
500 to 10000 Hz
25 Hz
30 to 80 dB SPL (75 dB SPL for 6kHz and 65 dB SPL for 8kHz to
10kHz)
1 dB
IOW Probe auto detection, auto calibrated
Minimum 2 seconds to unlimited time
24 bit, 5.38 Hz resolution
-30 to +30 dB SPL or off
Adjustable between 1 and 10 dB
Adjustable between 3 and 25 dB
256 points frequency response of the ear canal due to a click
stimulus.
4096 points frequency response
Applicable during testing
Stimulus level and type, Bar and Graph view
IMP, DPOAE and ABRIS capable
Replaceable probe tip
Ambient pressure.
Input/output for computer communication. Titan can be fully
operated from a PC. The measurements can then be followed
on the PC screen.
Data can be sent to and saved on the PC and stored in
OtoAccess™.
See separate section in Service Manual for programming
details.
Theoretically, an infinite amount of test results can be stored on
the PC. The Titan hand held unit is delivered with a 1 Gb
memory card, enough for storing more than a quarter of a
million tests.
Thermal (Bluetooth) printer with recording paper in rolls. Print on
command through Bluetooth communication and through serial
RS-232.
57.5 ± 0.5 mm on thermal printer
Printing time depends on the size of the used protocol. For 2
tympanograms and 8 reflexes the thermal printer uses
approximately
6s.
6x6x28cm/2.4x2.4x11inch
360g/0.8lbs
120 g
102, 68, 26 mm.
64 g
65, 35, 18 mm.
Page 16/21