Descargar Imprimir esta página

Memory Function - Gima TH889 Guia De Inicio Rapido

Termómetro auricular

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 18
7
4. Press "ON/MEM" button to power on. The thermometer is ready
for use after the ear icon stop flashing and two short beep sound.
5. Gently pull the ear back to straighten the ear canal and snugly po-
sition the probe into the ear canal, aiming towards the membrane of
the eardrum to obtain an accurate reading. (Fig.4-1)
6. Measuring the ear temperature: Use the index finger to trigger.
Press the "Scan" button until you hear a long beep. (Fig.4-2). After
each ear measurement, wait "
next measurement.
7. Power off: This device will automatically shut down after 1 minute
pending to extend battery life.
NOTE
a. Before the measurement, please stay in a stable environment for 5mins and avoid the exercise, bath
for 30mins.
b. It is recommended that you measure 3 times with the same ear. If the 3 measurements are different,
select the highest temperature.
c. To avoid the risk of cross contamination, please clean the probe according to "Cleaning and Storage"
section after each use.
d. The "Clinical Bias" is -0.2 ~ -0.4°C.
e. The "Limits of Agreement" is 0.58.
f. The "Repeatability" is 0.17°C.
Fever Indication:
If the thermometer detects a body temperature ≥37.5°C (or 99.5°F), three short beep sound will follow
one long beep sound to warn the user for potential fever.
Switching between Fahrenheit (°F) and Celsius (°C):
In "Power Off" mode, press and hold the "SCAN" button, then press the "ON/MEM" button for 3 seconds,
icon "°C" will be switched to icon "°F". You can also use the same process to change the LCD display
from °F to °C.

Memory Function:

When in power on, press the "ON/MEM" button to see the temperature stored. The thermometer pro-
vides 9 sets memory for the body temperature.
Cleaning and Storage:
The probe is the most delicate part of the thermometer. Use with care when cleaning the lens to avoid
damage.
** The probe cover is disposable: throw it away and replace it after each use to ensure an ac-
curate reading and avoid cross contamination.
Storage temperature Range: It should be stored at room temperature between -20~+50°C, RH ≤85%
Keep the unit dry and away from any liquids and direct sunlight. The Probe should not submerge into
any liquids.
** If device is accidentally used without probe cover, clean the probe as follows:
a. Please use the cotton swab with Alcohol (70% concentration) to clean the lens( on
the inside of the probe).
b. Allow the probe to fully dry for at least 1 minute.
NOTE: Please check the device if damaged once it falls. If you can't make sure of it,
please send the complete device to the nearest retailer for recalibration.
Holding the thermometer too long may cause a higher ambient temperature reading
of the probe.
This could make the body temperature measurement lower than usual.
" icon stop flashing to be ready for
ENGLISH
Figure 4-1
Figure 4-2

Publicidad

loading

Este manual también es adecuado para:

25580