Adjusting the plate pressure
Note
In order to increase the cutting
performance, the plate pressure may have
to be adjusted, see fold-out page!
The knurled screw 15 must only be
tightened so far that the upper cutting
plate 19 can still be moved by hand
from the side with the application of
only an average amount of pressure.
Adjusting the plate pressure with the
Torqui precision adjusting aid
Note
The manufacturer recommends using the
Torqui precision adjusting aid for optimal
adjustment of the plate pressure.
Insert the Torqui precision
adjustment aid into the socket hex of
the knurled screw and turn clockwise
until a click can be heard.
Note
Knurled screws of older clipping heads that
do not have a socket hex can be retrofitted
with the screw set GT141.
Disassembling the clipping head
Loosen knurled screw 15 with leaf
spring 16
Push upper cutting plate 19 under
pressure piece 14 out of the clipping
head
The clipping head is now disassembled!
Oiling the clipping head
 Disassemble the clipping head, see
chapter "Disassembling the clipping head".
 Using a soft cloth or brush, clean the
upper 19 and the lower 18 cutting plate.
While doing so, ensure that the gaps
between the teeth 17 are clean.
 Lightly oil the upper 19 and the lower 18
cutting plate using GT604 special oil for
clipping heads.
Grinding the cutting plate
Note:
For grinding of the cutting plate, please
contact the technical service department,
see "Maintenance".
Maintenance
To ensure reliable operation, Aesculap
recommends servicing the product at least
once a year.
Service addresses:
Albert Kerbl GmbH
Felizenzell 9
84428 Buchbach/Germany
Phone:
+49 (0) 8086933-100
Fax:
+49 (0) 8086933-500
E-Mail:
info@kerbl.com
www.kerbl.com
Other service addresses can be obtained
from the address indicated above.
19

Publicidad

loading

Este manual también es adecuado para:

Gt234-sr