Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic
material, the relative humidity should be at least 30%.
The main power quality should be similar to that of a typical commercial or
hospital environment.
The main power quality should be similar to that of a typical commercial or
hospital environment
The main power quality should be that of a typical commercial or hospital
environment. If the user of the equipment or system requires continued operation
during power main interruptions, it is recommended that the equipment or system
be powered from an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment.
Specifications & technical data
Product description: Digital automatic blood pressure monitor
Model: WPM05
Blood pressure measurement method: Cuff oscillometric method
Cuff inflation: Automatic inflation with air pump at 6 mmHg/s
Pressure sensor: Gauge sensor
Measurement range (pressure): 0 to 285 mmHg, DIA 40 to 130 mmHg, SYS 60 to
230 mmHg
Measurement range (pulse): 40 to 180 beats/min
Pressure sensor accuracy: Within +- 3 mmHg or 2% of reading
Clinical accuracy (blood pressure): BPM Connect performance in measure blood
pressure measurement was validated in a clinical trial. The results are within the margin
of acceptance defined by the internationally recognized evaluation standard
of blood pressure monitors ANSI/AAMI/ISO 81060-2:2013, EN ISO 81060-2:2014,
developed by the European Society of Hypertension, British Hypertension Society
and Association for the Advancement of Medical Instrumentation/American Heart
Association. For more information, go to: withings.com/support
Accuracy (pulse): Within +-5% of reading
Sensor: Semiconductor pressure sensor
Operating conditions: 5 to 40°C, up to 90% RH, atmospheric 86Kpa~106kpa,
altitude: 2000m
Storage and transport conditions: -20 to 70°C, up to 95% RH, atmospheric
86Kpa~106kpa, altitude: 2000m
Arm type: Use on left arm
Power source: 5V 1A
Weight: Approx. 245g
Accessories: Charging cable, Quick Start Guide, Product Guide
Product life: 3 years
Wireless transmission Wi-Fi and BLE
Note: Specifications are subject to change without prior notice or any obligation
on the parts of the manufacturer