RF Statement
Medical Electrical Equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in
the following. | Interference may occur in the vicinity of equipment marked with
the following symbol:
| Portable and mobile RF communication equipment (e.g.
cell phones) can affect Medical Electrical Equipment. | The use of accessories and
cables other than those specified may result in increased emissions or decreased
immunity | The device uses RF energy only for its internal function. Therefore,
itsRF emissions are very low and are not likely to cause any interference in nearby
electronic equipment | The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low voltage
power supply network that supplies buildings used for domestic purposes |
Portable and mobile RF communications equipment should be used no closer to
any part of the device, including cables, than the recommended separation dis-
tance calculated from the equation applicable to the frequency of the transmitter |
The device is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. Any other accessories, transducers and cables may
result in increased emissions or decreased immunity and EMC performance. |
The device should not be used adjacent to or stacked with other equipment and
that if adjacent or stacked use is necessary, which should be observed to verify
normal operation in the configuration in which it will be used. Medical Electrical
Equipment needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided in the following.
Portable and mobile RF communication equipment (e.g. cell phones) can affect
Medical Electrical Equipment. The use of accessories and cables other than those
specified may result in increased emissions or decreased immunity of the unit.
Guidance and manufacturer's declaration-electromagnetic emissions
This Wi-Fi Smart Blood Pressure Monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of the Wi-Fi Smart Blood
Pressure Monitor should ensure that it is used in such an environment.
Emissions test
Compliance
CE emissions
Group 1
CISPR11
RE emissions
Class B
CISPR11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
Not applicable
IEC 61000-3-3
8 – 9
Electromagnetic environment —
guidance
The Wi-Fi Smart Blood Pressure
Monitor uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
This Wi-Fi Smart Blood Pressure
Monitor is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
EN
FR
DE
SV
NL
FI
DA
IT
ES
CS
PL
PT
RO
HU
SK
ET
EL
IS
LV
LT
SL
TR
BG
HR
NO