Affinity2.0/Equinox2.0 – Instructions for Use - EN
Current Protocol is listed in the lower lef t-hand corner.
Upon perf orming a test, the system automatically jumps to the next one
in the test f low. The checkmarks indicates that a curve has been
measured.
Test protocols can be created and adjusted in the REM440 setup.
Colour on each test button indicates the colour selected f or each curve.
Start/Stop button initiates and ends the current test. Note that af ter
pressing START the text on the button will change to STOP.
The Graph shows measured REM curves. The X axis shows the
f requency, and the Y axis shows the intensity of the test signal.
Gain/Response View allows f or toggling between viewing the
curve as a gain or response curve. Note that this option is not
active f or REIG.
Measurement Type is indicated above the graph, with a right/lef t
indication. In this example the REIG is displayed f or the right ear.
Change the Input Level using the slider on the right-hand side.
Scroll Graph Up/Down on the lef t-hand side allows f or scrolling
the graph up or down ensuring that the curve is always visible in
the middle of the screen.
Fitting Prescription and related details can be adjusted on the
right-hand side of the screen. Select your pref erred f itting
prescription in the upper dropdown list.
Choose between Berger, DSLv.5.0, Half Gain, NAL-NL1, NAL-
NL2, NAL-R, NAL-RP, POGO1, POGO2, Third Gain, or 'Custom'
if you have edited your target with the Edit f eature.
Target shown will be calculated based on the selected f itting
prescription and the audiogram and can be shown as REIG
and/or REAR targets. If no audiogram has been entered into
the audiogram screen, no targets will be displayed.
Note that f itting prescription settings (such as Age and Client
type) will dif fer depending on what f itting prescription is selected.
Measurement Details of the selected curve are displayed as a
table on the right-hand side of the screen.
Page 39