1
1. Open the module cover.
2. Squeeze the two external locking lugs in and withdraw the
sealing insert downwards.
3. With the Pac/X-am 125 module: Remove the hoses from
the sealing insert (1).
With the X-am 8000 module: Loosen the screws at the
hose coupling and remove the hose coupling with the
hoses. Check that the seals of the hose coupling are intact,
replace the hose coupling (2) if required.
4. Replace the sealing insert.
5. With the Pac/X-am 125 module: Fit the hoses to the new
sealing insert (note the arrows on the sealing insert and the
hose).
With the X-am 8000 module: Fit the hose coupling with the
hoses on the new sealing insert and tighten the two
screws.
6. Squeeze the external locking lugs in and insert the sealing
insert into the module cover. The locking lugs must click
into place.
7. Check that the sealing insert is fitted correctly in the module
cover.
6.7
Changing the fresh air filter
NOTICE
i
i
With regular use and depending on the conditions of
use, the fresh air filter should be changed typically
every 2 months.
1. Unscrew the old fresh air filter.
2. Screw in the new fresh air filter.
6.8
Calibrating the touchscreen
1. At system start hold the function key until the calibration
screen is displayed.
2. Press each of the 5 position markers displayed in succession.
Dräger X-dock 5300/6300/6600
3
2
10333286.eps
6.9
Cleaning
CAUTION
!
Impairment and wear of the system!
Heavy contamination or dust accumulation may impair
the function of the system and lead to increased wear
of individual system components.
Clean the system at regular intervals.
Abrasive
cleaning
cleaning agents and cleaning solvents can destroy the
fresh air filter.
The device does not need any special care.
If very dirty, the equipment can be carefully wiped down
with a damp cloth.
Carefully dry the device using a cloth.
7
Disposal
This product must not be disposed of as municipal
waste. This is indicated by the adjacent icon.
You can return this product to Dräger free of charge.
For information please contact the national marketing
organisations and Dräger.
8
Technical data
Dimensions (H x W x D):
Master
approx. 3.54 x 5.70 x 9.84 inches
(120 x 130 x 250 mm)
Pac / X-am 125
approx. 3.54 x 5.70 x 9.84 inches
module
(90 x 145 x 250 mm)
X-am 8000 module
122 x 145 x 250 mm
Weight:
approx. 4.02 pounds
Master
(48.23 ounces; 1500 g)
Pac / X-am 125
approx. 2.57 pounds
module
(30.86 ounces; 960 g)
X-am 8000 module
approx. 1225 g
X-am 8000+ module approx. 1445 g
Ambient conditions:
During operation
32 °F to +104 °F (0 °C to +40 °C)
During storage
-4 °F to +122 °F (-20 °C to +50 °C)
700 to 1300 hPa
max. 95 % relative humidity
During charging
0 °C to +35 °C (outside of this
(X-am 8000+
temperature range charging is possible
module only)
only to a limited extent)
Disposal
implements
(brushes,
etc.),
31