Operating messages – Error codes
Operating messages
Hı
Measuring range exceeded or excessive turbidity.
Result below the lowest limit of the measuring range.
Lo
Replace batteries, no further tests possible.
btLo
Battery capacity is too low for the display backlight;
measurement is still possible.
Store
Date
A user calibrated method is indicated by a "Cal" symbol while the test
Print
Time
Cal
RESULT
result is displayed. (see "Factory calibration reset").
Error codes
Light absorption too great. Reasons: e.g. dirty optics.
E27 / E28 / E29
Calibration factor "out of range"
E 10 / E 11
Too much light reaching the detector.
E 20 /E 21
E23 / E24 / E25
Too much light reaching the detector.
E 22
Battery capacity was too low during measurement. Change battery.
E 70
Mo 1 (Powder Pack, LR):
E 71
Mo 1 (Powder Pack, LR):
E 72
Mo 2 (Powder Pack, HR):
E 73
Mo 2 (Powder Pack, HR):
E 74
Mo 3 (Tablet):
E 75
Mo 3 (Tablet):
Instrument accessory information
IRiM
24 mm Sample Cells, 12 pack
Light shielding rings
Cleaning cloth for sample cells
Sample cell rack
Mixing Cylinder, 25 ml glass w/ stopper
SC450_V1_04/2011
Factory calibration incorrect / erased
User calibration incorrect / erased
Factory calibration incorrect / erased
User calibration incorrect / erased
Factory calibration incorrect / erased
User calibration incorrect / erased