Window Meanings
N.
Read/
Data
Description
Write
Type
000
R/W
L
Start/Stop
(in remote mode
the window is a
read only)
008
R/W
L
Remote
(default) or Serial
configuration
100
R/W
L
Soft Start
(write only in Stop
condition)
101
R/W
N
Set Point type
102
R/W
N
Set Point
threshold
(expressed in Hz,
mA or s)
103
R/W
N
Set Point delay:
time between the
pump start and
the set point
check (seconds)
104
R/W
L
Set Point signal
activation type:
the signal can be
"high level active"
or "low level ac-
tive"
105
R/W
N
Set point
hysteresis
(in % of threshold)
106
R/W
L
Water cooling
107
R/W
L
Active Stop
(write only in stop)
108
R/W
N
Baud rate
109
W
L
Pump life/ cycle
time/ cycle num-
ber reset
110
R/W
L
Interlock type
(default = 1)
111
R/W
L
Analog output
type: output volt-
age signal propor-
tional to frequency
or power
120
R/W
N
Rotational fre-
quency setting
(Hz)
All manuals and user guides at all-guides.com
Admitted
Values
Start = 1
Stop = 0
Remote = 1
Serial = 0
(default = 1)
YES = 1
NO = 0
(default = 1 the first
time and 0 the fol-
lowing)
0 = Frequency
1 = Current
2 = Time
(default = 0)
(default = 867)
0 to 99999
(default = 0)
0 = high level ac-
tive
1 = low level active
(default = 0)
0 to 100
(default = 2)
0 = NO
1 = YES
0 = NO
1 = YES
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
150 to 963
(default = 963)
35
TECHNICAL INFORMATION
N.
Read/
Data
Description
Write
Type
121
R/W
N
Maximum rota-
tional frequency in
Hz (active only in
Stop condition)
122
R/W
L
Set vent valve
on/off
(on = closed)
123
Reserved to Varian service
124
125
R/W
L
Set the vent valve
operation
126
R/W
N
Vent valve opening
delay
(expressed in 0.2
sec)
130
Reserved to Varian service
200
R
N
Pump
current in mA dc
201
R
N
Pump voltage in
Vdc
202
R
N
Pump power in W
(pump current x
pump voltage duty
cycle)
203
R
N
Driving frequency
in Hz
204
R
N
Pump tempera-
ture in °C
205
R
N
Pump status
206
R
N
Error code
210
R
N
Actual rotation
(see
speed in Hz
note 3.)
300
R
N
Cycle time in
minutes (zeroed
by the reset com-
mand)
301
R
N
Cycle number
(zeroed by the re-
set command)
302
R
N
Pump life in hours
(zeroed by the re-
set command)
320
Reserved to Varian service
to
399
Admitted
Values
150 to 963
(default = 963)
On = 1
Off = 0
(default = 1)
Automatic = 0 (see
note 1.)
On command = 1
(see note 2.)
0 to 65535
(corresponding to 0
to 13107 sec)
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description: see
the following figure
100 to 963
9
0 to 4x10
9
0 to 4x10
9
0 to 4x10
87-900-946-01(B)