Aging; Maintenance - wtw F 500 DIN Instrucciones De Operación

Publicidad

Idiomas disponibles

Idiomas disponibles

F 500 / F 800
Opening the
combination elec-
trode
ba15358e03
03/2015

Aging

Please note that every electrode undergoes a natural aging process. The re-
sponse time increases and the slope decreases with the age of the electrode.
The following factors shorten the lifetime considerably:
 Unsuitable storing conditions (see section S
 Special measuring media (e.g. organic solutions)
 High temperatures
 High changes in temperature
The warranty does not cover damage caused by measuring conditions and
mechanical damage.

Maintenance

 Combination electrodes:
Refill any used up bridge electrolyte.
 In the case of increased response times of the electrode perform the fol-
lowing activities:
– In the case of organic coatings clean the membrane with isopropanol
and a soft paper towel.
– Then recondition the electrode (see section C
,
CALIBRATION
MEASUREMENT
Clean the inside of the combination electrode. To do so, open the
combination electrode as follows:
1
Unscrew the knurled nut from the connection head.
2
Position the connection cable upward in a straight line with the con-
nection head.
3
Unscrew the connection head.
4
Push the connection head and pressure spring over the connection
cable.
5
Push the inner body out of the shaft.
NOTE
Never pull the connection cable of the electrode.
The cable might be damaged.
TORAGE
ONDITIONING
).
Aging
)
,
23