Storage

Between two
measurements
Overnight
Longer than over-
night
24
Storage
Condition the electrode in standard solution.
F 500: Condition the electrode in standard solution.
F 800: Condition the combination electrode in standard solution with the
filling opening open. To avoid a contamination of the bridge electrolyte with
standard solution, the level of the bridge electrolyte must be clearly above the
level of the conditioning solution.
For the following measurement, exchange the bridge electrolyte. To do so,
press the shaft of the combination electrode against the connection head and
let the bridge electrolyte drain off. Then fill new bridge electrolyte into the
filling opening.
F 500: Rinse the electrode with deionized water, then dab it dry with a clean
paper towel and put on the protective cap. Store the electrode in a dry place.
F 800: Remove the bridge electrolyte, then rinse the combination electrode
with deionized water and dab it dry with a clean paper towel. Store the combi-
nation electrode in the storage vessel in a dry place. For the following
measurement, prepare the combination electrode as described in section
C
.
OMMISSIONING
Store the reference electrode according to the instructions in its operating
manual.

Recommended accessories

Description
Reference electrode for electrode F 500
Standard solution 10 g/l F
Bridge electrolyte
TISAB sample conditioning solution
*
Pin plug
**
Banana plug
-
F 500 / F 800
Model
Order no.
R 503/P*
106570
R 503/D**
106571
ES/F
120160
ELY/BR/503
106 575
TISAB
140100
ba15358e03
03/2015