OFF button and the main unit will produce a "Bi" sound, the curing light radiates blue
light and starts working according to the set modes meanwhile. lt starts counting down
and will produce a "Bi", it stops working when counting down to 0. At the end of the
work, the digital display returns to the set work mode.
– During operation, the blue light can be stopped by press the power button at any time.
– After a working cycle, the operator can press the ON/OFF switch to start another working
cycle. Stop operating if the equipment starts to get noticeably hot, do not restart until
the equipment has cooled down. We recommend not to perform more than 10 working
cycles immediately after the other.
– Low power detective circuit is fi xed inside of the main unit when low power is detected,
the display screen of main unit will wink, please charge in time.
– When the battery needs to be charged, connect the plug of the adapter into the ac
100 V ~ 240 V power supply. Then connect the output plug of the adapter to the DC
5.0 V input plug of the charger station. Put the main unit to the charging point of the
charger station, the display screen turned to scrolling display, and the curing lights
starts charging.
– This equipment will turn o automatically if there is no any action within 7 minutes; You
can turn it on by press any button.
– The depth of solidifi cation of composite is no less than 4mm per 10 second.
6. Clean and Disinfection
ln order to prevent cross-infection, it's commended to do following process before and
after each operation.
1. Turn o the device and take the main unit from the charger.
2. Take o the cover sleeve after operation.
3. Wipe the main unit with a soft cloth moistened with purifi ed water.
4. Let the device air dry.
7. Daily maintenance
– This equipment does not include the self-maintainable spare parts.
– The maintenance of this equipment should be taken by the appointed professional or
special repair shop.
– Users can change the lithium battery. Please use accessories which are designed and
supplied by our company, contact with the local dealer or our company if you want to
buy. Accessories designed and supplied by other manufacturers may cause potential
hazards to the unit or other damage.
– The accessories of the product should be cleaned with clean water or neutral sterilized li-
quid. Do not soak. Do not use highly volatile and di uent solvent to clean this equipment,
which can cause the signs on the control panel to fade.
– Please clean the resin remained on the top of the main unit after using to
the
life-spanor
solidifi ed e ect.
– The device cannot be maintained during operation, and it is suggested to maintain it
once a month.
avoid infecting
15