10. Documentación de mantenimiento/Certificado del dispositivo
SN:
CPU-Monitor SN:
Software Version:
Blender SN:
Front panel:
Test equipment:
Test of oxygen calibration:
21% ±3%
60% ± 3%
Test of pressure calibration:
0 cmH2O ± 0,5
Overpressure valve:
Alarm test:
Dissconnection alarm
(P<1,5cmH20 ~ 5sec)
nCPAP low alarm (P< CPAP
low alarm ~ 20sec)
Audio alarm (dis)able
Oxygen high low / alarm
Gas supply pressure Air/O
2
Electric check:
Battery check
Device check without
power supply
Power supply
medin Medical Innovations GmbH
Adam-Geisler-Str. 1
D-82140 Olching
Gerätepass/Device Passport
®
medinSINDI
Akku-SN:
Oxygen cell SN:
Gas connection system:
sindi
Oxygen measurement
(IF)
Oxi Quant MC
(Envitec)
40% ±3%
ok /
nok
ok /
nok
100% ±3%
10cm H2O ± 0,5
ok /
nok
ok /
nok
High pressure alarm
ok /
nok
P> 10cmH2O ~5sec
ok /
nok
nCPAP high alarm (P>CPAP
high alarm ~5sec)
Audio alarm off / on
ok /
nok
ok /
nok
ok /
nok
Charge
ok /
nok
ok /
nok
ok /
nok /
LOT No:
at customer
+49 8142 448 460
info@medingmbh.com
www.medingmbh.com
REF 1080
Flow analyser PF-302
LOW (IMT)
Flow analyser PF-300
(IMT)
OP_medinSINDI ES
Rev 14
Stand 30.10.2017
DISS
NIST
Mass flowmeter
Model 4143 (TSI)
Manometer 0-25 cm WS
(Suchy)
ok /
nok
ok /
nok
ok /
nok
ok /
nok
ok /
nok
ok /
nok
ok /
nok
SN:
34 von 35