Descargar Imprimir esta página

Mounting / Exchanging The Saw Blade; Releasing The Retaining Screw / Removing The Old Saw Blade - HEBU medical GOLD II Autopsie Instrucciones Para El Uso Y El Servicio

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Oscillo-Saw GOLD II Autopsy

5 Mounting / exchanging the saw blade

If applicable use thick gloves to prevent injury when handling the saw
Only ever connect the HEBU oscillating cast cutting saw to the mains

5.1 Releasing the retaining screw / removing the old saw blade

1. Place the cast cutting saw on a stable underlay (table).
2. Apply one fork wrench at the drive shaft behind the saw blade,
and the other one at the front at the retaining screw (see fig. 1).
3. Hold the rear wrench firmly and use it to fix the cast cutting saw.
4. Release the retaining screw using the front wrench.
5. Remove the retaining screw and the fixing ring.
Ensure that you do not lose any components.
6. If applicable, remove the old saw blade.
Danger of injury due to unintentional activation.
Pull out the plug before changing the saw blade.
blade.
when it is switched off.
Fig. 1:
Releasing the retaining screw
EN
23

Publicidad

loading

Este manual también es adecuado para:

Hb 8891-aHb 8891-bHb 8891-gHb 8891-sHb 8891-v