Oscillo-Saw GOLD II Autopsy

6.2 For the autopsy

The oscillating saw is used in the autopsy for cutting bones. The guiding of the saw is described
in 4.1.
In pathological use, hard bandages can be also opened with the Autopsy saw if necessary.
Please note also the additional information in 4.1!

7 Cleaning and maintenance

Danger of injury through unintentional
activation.
Pull out the plug before any cleaning or main-
tenance work.
Wipe disinfection is suitable for motor housing, cable, power plug and all accessories, such as saw
blades and an open-end wrench. For this purpose, we recommend
products with a pH-value within 9-10, e.g. MediClean forte from Dr.
Weigert.
It is important to ensure that the device is disconnected from the
power supply and the disinfecting liquid does not enter the interior of
the motor.

7.1 The use of the cast saw.

The following measures should be performed once a week
Fig. 5
or more often if the saw is used frequently:
Clean the cooling slots (see fig. 5) to ensure sufficient ventilation for cooling the motor.Clean the
motor compartment from the outside with a jet of dry compressed air.

7.2 The use of the autopsy saw:

The saw should be cleaned after every use.
An efficient cleaning can only be made if you remove the
saw blade. (Compare 3.1 Releasing the retaining screw/
removing the old saw blade). Remove the two head screw.
The retaining screws, fixation ring and saw blade have to be
revomed. Take off the saw head (compare fig. 10).
Fig. 10
Warning!
Please use thick gloves to prevent injury
when handling the saw blade
For the machine cleaning the saw head should lie down
horizontal- compare fig. 11. The machine cleaning should be
effected with a standard cleaning program. The cleaning agent
must have a pH-value of <10, otherwise the surface coating will
be damaged.
26