Descargar Imprimir esta página

Braun ExactFit 1 Manual Del Usuario página 9

Ocultar thumbs Ver también para ExactFit 1:

Publicidad

Idiomas disponibles

Idiomas disponibles

Calibration
This product has been calibrated at the time of manufacture. If used
according to the instructions for use, periodic recalibration is not
required. If at any time you question the measuring accuracy, please
contact our service representative (see last page for contact information).
What to do if ...
Error
«Err 1»
«Err 2»
«Err 3»
«Err 5»
«HI»
«Lo»
Description
Batteries are low
Batteries are fully
discharged
Signal too weak
Error signal
No pressure in the
cuff
Abnormal result
Pulse or cuff
pressure
too high
Pulse too low
Potential cause and remedy
Get ready to replace batteries soon with
new ones.
Insert new batteries or check that
batteries are correctly inserted.
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.
During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle
tension. Repeat the measurement,
keeping the arm still.
An adequate pressure cannot be
generated in the cuff. A leak may have
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
The measuring signals are inaccurate
and no result can therefore be
displayed. Read through the checklist
for performing reliable measurements
and then repeat the measurement.
The pressure in the cuff is too high (over
300 mmHg) OR the pulse is too high
(over 200 beats per minute). Relax for
5 minutes and repeat the measurement.
The pulse is too low (less than 40 beats
per minute). Repeat the measurement.
7

Publicidad

loading

Este manual también es adecuado para:

Bua5000