FAQS
Fault
Start failure
Infrared image is
not clear
Visible light image
is not clear
Inaccurate
temperature
measurement
Cause
Low battery
The plug of the
external power supply
is not plugged in place
Battery life has expired Replace the battery with a
The lens is covered
with water vapor or
contaminated
The environment is
too dark
Water vapor or
contamination in the
front end of visible
light
The parameters
related to temperature
measurement are not
set correctly
Measure temperature
immediately after
startup
No calibration for a
long time
20
Measures
Start the device after charging
for 10 minutes
Unplug, reconnect and push it
into place
new one
Clean the lens with dedicated
device
Take proper lighting measures
Clean the visible light front
end with dedicated device
Change the parameter
settings or restore the default
parameter values directly
To ensure the accuracy of
temperature measurement,
we recommend to start
temperature measurement
after 5 ~ 10 minutes when the
device is turned on
For accurate temperature
measurement results, we
recommend to calibrate the
thermographic camera once
a year