AC Current (Average sensing RMS indicating)
Resolution
Range
400µV
4mA
40mA
400mA
2000mA
20A**
Overload Protection:
2A/600V fuse on mA inputs.
20A/600V fuse on 20A inputs.
** 20A for 30 seconds maximum.
Resistance
Resolution
Range
400Ω
4kΩ
40kΩ
400kΩ
4MΩ
40MΩ
Overload Protection: 500V
Continuity Test
Range
Threshold
Less than
4V
Overload Protection: 500VDC or RMS AC
Diode Test
Range
Resolution
4V
Overload Protection: 500VDC or RMS AC
Logic Test
Thresholds
Logic 1
Logic 0
(Hi)
2.8V±0.8V 0.8V±0.5V
10/06
Accuracy (50Hz to
µ
0.1
A
1µA
±(1,5% rdg + 1 d)
10µA
100µA
1mA
±(2,0% rdg + 1 d)
1mA
±(3.5% rdg + 3 d)
0,1 Ω
±(1.0% rdg + 4 d)
1 Ω
±(0.75% rdg + 4 d)
10 Ω
±(0.75% rdg + 4 d)
100 Ω
±(0.75% rdg + 4 d)
1k Ω
±(0.75% rdg + 4 d)
10k Ω
±(2.0% rdg + 5 d)
DC
Audible
Response Time
100Ω
Accuracy
±(1.5% rdg
1mV
+ 3d)
Pulse Rise
(máx)
(Lo)
10µSec
500Hz)
Accuracy
or RMS AC
Approx. 100ms
Test
Current
1.0mA
Pulse Rep
(máx)
1Mpps
- 3 -
Burden Voltage
600mV
600mV
600mV
600mV
800mV
800mV
Open Circuit
Volts
3.2Vdc
0,6Vdc
0,6Vdc
0,6Vdc
0,6Vdc
0,6Vdc
Open Circuit
Volts
3.2Vdc typical
Open Circuit
Volts
3.2Vdc
typical
Pulse
Width
(máx)
25ns