9.6
Use of additional lens
•
Additional lenses can be used on models SLX-2 / SLX-3 /
SLX-4 / SLX-5.
1. Screw the desired additional lens on the microscope body.
(Fig. 14)
•
Each additional lens has a specific Working Distance (see
table below).
•
(SLX-2 / SLX-3 only): The stroke of the focus adapter could
not compensate the different working distances of the sever-
al additional lenses.
Total magnification used can be calculated as:
Eyepiece magnification * Zoom magnification * Objective lens magnification.
Eyepiece
Field number
Objective
0,5X (W.D.165 mm)
0.75X (W.D. 114 mm)
5.25X-33.75X
1X (W.D. 100 mm)
1.5X (W.D. 47 mm)
10.50X-67.5X
9.6.1
Use of 0.5X additional lens (ST-085.1)
0.5X additional lens has a Working Distance of 165 mm and
therefore the total stroke of the focus pillar cannot compensate.
In order to let the 0.5X additional lens working properly (when
used with SLX-2 / SLX-3 models), proceed as follows:
1. Screw the additional lens on the microscope body as already
described in chapter 9.6.
2. Uninstall the microscope body from the head holder.
3. Install the provided spacer ① in the head holder and screw
the fixing knob ②.(Fig. 24)
4. Re-install the microscope body onto the spacer and screw
the fixing knob ③.(Fig. 25)
5. Begin working normally.
10X
21
Total mag.
F.O.V. (mm)
3.5X-22.5X
59.99-9.32
40-6.21
7X-45X
29.99-4.66
20-3.10
15X
15
Total mag.
F.O.V. (mm)
5.25X-33.75X
28.57-4.44
7.87X-50.62X
19.06-2.96
10.5X-67.5X
10.95-2.22
15.75X-101.25X
9.52-1.48
①
②
Page 16
F ig. 23
F
ig. 23
20X
10
Total mag.
F.O.V. (mm)
7X-45X
14.28-2.22
10.50X-67.5X
9.52-1.48
14X-90X
7.14-1.11
21X-135X
4.76-0.74
F ig. 24
F
ig. 24
F ig. 25
F
ig. 25
③