Original instructions
Parameters:
Display
Description
Zero calibration of the ambient pressure
S 0
(0.0kPa - 0.00"Hg)
S P
To change the Set Point
S U
To change the measurement unit
Enable/disable the control and the errors on
O 1
output signal 20-4mA
Enable/disable the control and the errors on
O 2
lamp output
ESC
To esc from programming mode
SETTINGS
Zero calibration: at the first installation, the
DVG500 EVO require a manual calibration of the
ambient pressure. Disconnect the tube from the
vacuum line.
Enter in programming mode and enter into "S 0"
parameter. The display shows the vacuum level.
Keep the button pressed for 3 seconds to set this
value as the new 0.0kPa (0.0 PSI)
Set Point: to set a vacuum level as a Set Point,
connect the tube to the vacuum line with the
desired vacuum level. Enter in programming
mode and enter into "S P" parameter. The display
shows the vacuum level. Keep the button pressed
for 3 seconds to set this value as the new Set
Point.
Unit of measure: To modify the unit of
measurement, enter in programming mode and
enter into "S U" parameter. The display shows the
actual unit. Keep the button pressed for 3 seconds
to change the unit of measurement.
If the device is used as a vacuum gauge and it's
not connected to the inverter, it's necessary to
This manual is the property of InterPuls S.p.A. - any copying, even partial, is strictly prohibited.
Default
//
//
kPa
On
Off
//
disable the corresponding output to avoid errors of
"missing connection" . Enter in programming
mode and enter into "O 1" parameter. The display
shows the actual value. If it's "On", keep the
button pressed for 3 seconds to set it on "Off".
If the device is connected to the alarm light
(optional),
it's
necessary
corresponding output. Enter in programming
mode and enter into "O 2" parameter. The display
shows the actual value. If it's "Off", keep the
button pressed for 3 seconds to set it on "On".
Since the device is a EEE product, it
must not be disposed of as urban
waste, but must be managed by a
separate waste connection centre
through specific WEEE waste
collection systems.
Proper disposal of the product will
contribute to protecting the
environment.
5
5100067_08.22_EN
Possible values
-100÷0kPa
-100÷0kPa
kPa / "Hg
On / Off
On / Off
//
to
enable
the
EN