OXY 11
Calibration and
storage vessel
ba75842e04
05/2021
4.2
Measuring
Note the required minimum depth of immersion and minimum approach flow
(see chapter 7 T
ECHNICAL DATA
The minimum flow can be provided in different ways, e. g.:
The flow of the water to be measured is sufficient (aeration tank, water pipe,
stream)
Slowly pull the sensor through the water by hand (lake, container), or
Use a flow aid, e.g. stirrer.
4.3
Storage
Always store the sensor in the calibration and storage vessel at a temperature
in the range 0 ... +50 °C (32 ... 122 °F). Make sure that the sponge in the
calibration and storage vessel is always moist.
1
2
Moisten the sponge:
Pull out the insert (1).
Take out the sponge (2), wet it, then slightly squeeze it out.
Put in the sponge and install the insert in the calibration and storage vessel
again.
).
EN
English
7