17. Align the "D"-shaped part of the battery pack with the pins on the shaft.
18. Slowly move the assembly onto the housing. Push gently to connect the battery assembly with
the circuit board in the pressure housing.
19. Use a 7/64" hex key to tighten the captured screw on the yellow battery cover onto the shaft in
the pressure housing.
20. Attach the Molex connector on the end flange to the connector in the pressure housing.
21. Examine the O-rings on the end flange. They must be pristine, with no lint or scratches or chips.
• Apply a small quantity of Parker Super O Lube
22. Align the end flange holes with the holes in the pressure housing.
23. Carefully push the end flange into the pressure housing.
It may help to rotate the end flange so that the wires do not bend too much.
24. If necessary, use a 9/64" hex key to remove the two screws from the end flange.
25. Use a 9/64" hex key to install the two screws into the pressure housing of the sensor again.
Section 4 Antifouling devices
AF24173 anti-fouling devices contain bis(tributyltin) oxide. Wear rubber or latex gloves and eye protection to
replace these devices on the sensor if it is so equipped. Wash hands with soap and water when finished.
Read the precautions on the product label.
It is a violation of US federal law to use this product in a manner that is inconsistent with its label.
8 English
®
to any new O-rings.
C A U T I O N