cleaning and maintenance
5
Nebulizer Cleaning (JetNeb Kit)
All parts of the nebulizer, except tubing, should be cleaned according to the following
instructions. Your physician and/or drug provider may specify a certain cleaning procedure.
If so, follow their recommendations.
WARNING: Clean nebulizer after each use. Risk of bacteria
growth, infection, illness from contamination due to dirty nebulizer.
Cleaning:
1.
Ensure that the power cord is disconnected from the power outlet.
2.
Disconnect nebulizer from tubing and compressor. Do not wash or
submerge the air tubing.
3.
Disassemble all parts of nebulizer by twisting the top and bottom,
separating the pieces,
4.
Clean nebulizer in warm soapy water.
5.
Rinse all parts with clear water to remove any residue.
6.
Allow the nebulizer and all parts to air dry completely and store in a clean plastic
bag or container.
The 18260 JetNeb Nebulizer is intended for single patient
use and should not be shared with other people. This unit
should be replaced every thirty days.
Use and Cleaning of NEBKIT 500: refer to the user manual of NEBKIT 500 which is included
in the accessory pack of NEBKIT 500 independently
Compressor Cleaning
1.
With power switch in the "Off" position, unplug power cord from wall outlet.
2.
Wipe outside of the compressor cabinet with a clean damp cloth to
keep dust-free.
CAUTION: do not use any powdered cleaners or soap
pads which may damage the finish.
WARNING: Do not submerge the compressor in water;
doing so will result in compressor damage.
Filter Change
.............................................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL MQ5800, MQ5900, MQ5900P
.............................................................................................................................
WARNING!
CAUTION
CAUTION
WARNING!
11