Guidance and manufacturer's declaration - electromagnetic Immunity
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of
the Device should assure that it is used in such an environment
Immunity
IEC 60601
Test
Test level
Conduced RF
150KHz to
IEC61000-4-6
80MHz:
3Vrms
6Vrms (in ISM
and amateur
radio bands)
80% Am at
1kHz
Radiated RF
10V/m
IEC61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the [XXXXX] is used exceeds
the applicable RF compliance level above, the [XXXXX] should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the [XXXXX].
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
.............................................................................................................................
DRIVE COMPRESSOR/NEBULIZER MODEL MQ5800, MQ5900, MQ5900P
.............................................................................................................................
Compliance
Electromagnetic environment - guidance
level
150KHz to
Portable and mobile RF communications equipment should
80MHz:
be used no closer to any part of the Device ,including
3Vrms
cables, than the recommended separation distance calcu-
6Vrms (in ISM
lated from the equation appropriate for the frequency of the
and amateur
transmitter. Recommended separation
radio bands)
distances:
80% Am at 1kHz
d = 1.2√P;
d = 2√P
10V/m
80MHz to
800MHz:
d=1.2
800MHz to
2.7 GHz:
d = 2.3√P
Where P is the maximum output power
rating of the transmitter in watts (W) ac-
cording to the transmitter manufacturer,
d is the recommended separation dis-
tance in meters (m) Field strengths from
fixed RF transmitters, as determined by
an electromagnetic site surveya ,should
be less than the compliance level in each
frequency range b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
17