β-Ketone: Accuracy
The accuracy of the the GlucoMen
assessed by comparing blood β-ketone results obtained by patients
with those obtained using a Randox Monaco, a laboratory instrument.
Slope
Y-intercept
Correlation coe cient (r)
Number of subjects
Range tested
/ 10.7
β-Ketone: Precision
The precision studies were performed in a laboratory using GlucoMen
Day METER 2K system.
Within Run Precision
*Bloodav 0.5 mmol/L
*Bloodav 1.3 mmol/L
*Bloodav 3.5 mmol/L
*Bloodav 5.6 mmol/L
*Bloodav 7.3 mmol/L
Between Run Precision
*Controlav 0.80 mmol/L
*Controlav 2.65 mmol/L
*Controlav 4.74 mmol/L
This study shows that there could be variation of up to 4.5 %.
33
Day METER 2K system was
®
0.9922
0.03908 mmol/L
0.994
200
0.01-7.66 mmol/L
SD = 0.039 mmol/L
SD = 0.061 mmol/L
SD = 0.017 mmol/L
®
CV = 3.9 %
CV = 3.7 %
CV = 4.5 %
CV = 2.7 %
CV = 3.2 %