3)
Two fuses are mounted on the rear of the IEC
connector (both 'Live' and 'Neutral' mains
supply leads are fused). These fuses can be
removed and replaced by hand. ALWAYS
replace the fuse by ones of the correct values
shown below:
For 230V, 50Hz - T1A
For 120V, 60Hz - T1.6A
4)
Refit the base and four securing screws.
5)
It is advisable to perform a 'PAT' (Portable
Appliance Test) electrical safety test, or
equivalent, before the unit is put back into
operation.
For a comprehensive list of parts and a Service
Manual required by service engineers conducting
internal repairs, please contact the Technical Service
Department of Bibby Scientific Ltd. quoting both the
model and serial number.
Only spare parts supplied or specified by Bibby Scientific
Ltd. or its agents should be used. Fitting of non-
approved parts may affect the performance and safety
features designed into the instrument. If in any doubt,
please contact the Technical Service Department of
Bibby Scientific Ltd. or the point of sale.
Bibby Scientific Ltd.
Stone, Staffordshire ST15 0SA
United Kingdom
Tel: +44 (0)1785 812121
Fax: +44 (0)1785 813748
e-mail info@bibby-scientific.com
www.bibby-scientific.com

Technical Specification

Number of samples
Temperature range
Temperature resolution
Temperature accuracy
Display
Sensor
Ramp rates
Dimensions (WxDxH)
Net weight
Warranty
Bibby Scientific Ltd. warrants this instrument to be free
from defects in material and workmanship, when used
under normal laboratory conditions, for a period of
three (3) years. In the event of a justified claim Bibby
Scientific will replace any defective component or
replace the unit free of charge. This warranty does NOT
apply if damage is caused by fire, accident, misuse,
neglect, incorrect adjustment or repair, damage caused
by incorrect installation, adaptation, modification,
fitting of non-approved parts or repair by unauthorised
personnel.
SMP10
Two
Ambient to 300°C
1°C
±1°C
Three digit LED
PT100
20°C / minute to plateau
2°C /minute to melt
160 x 220 x 170 mm
1.8 Kg
SMP20
Two
Ambient to 300°C
0.1ºC
±1°C
Four digit LED
PT100
20°C / minute to plateau
Between 1 and 10ºC/min
160 x 220 x 170 mm
1.8 Kg