The battery charge status is indicated by the battery icon on the display. The battery icon will
indicate when the battery charge is full, partial, low, empty or charging. The empty battery icon
will flash to indicate that meter should be connected to AC power source. If the meter continues
to be used at low battery power without connecting it an AC power source, the meter will go into
an auto-shutdown mode. In this mode the meter will be locked until meter is connected to an AC
source and the battery is charged to a sufficient voltage.
While charging, the charging battery icon will be displayed. The meter should
remain plugged in until the battery is fully charged. When the battery is completely
charged, the charging icon will change to the full battery icon.
SETTINGS
Tap
to enter the SETTINGS menu. Tap
Brightness The brightness level of the display can be adjusted from 00 to 10. Tap
adjust the brightness. Tap
Date/Time The Year, Month, Day, Format, Hour, Minute, AM/PM can be set. Tap
adjust the displayed value. Tap
(minutes for 24 hour format, AM/PM for 12 hour format) tap
Tap
to exit to the Settings menu at any time.
Set Language There are ten language options – English, French, Italian, Portuguese, Spanish,
Chinese, German, Dutch, Swedish, Turkish. Tap selection. Tap
Calibration Tap to run an angle calibration to evaluate the alignment of the hub and disk. Tap
to exit to the Settings menu.
Power Options There are three power options: Auto Dim Time, Auto Off Time, and Power. Tap the
options then tap a selection. Tap selection. Tap
Bluetooth There are two Bluetooth options: Bluetooth Enabled and Fast Printer Connect that are
used to transfer and print test results. Tap
Market Type Tap to select the market type. Select Pool/Spa. Tap
Other Settings About...lists the Serial Number, Firmware Version, Bluetooth MAC address,
Bluetooth Version and Test Count. The Test Count shows the number of complete tests that have
been performed over the lifetime of the meter. Tap
Ranges Enabled allows the option of having test results that are out of range of the reagent
system displayed in red. The default setting is off.
When Show Market Type is selected, the market type will be displayed on the Testing Screen. The
default setting is OFF. Tap
RANGES
Test Factor
Alkalinity, Total
Biguanide
Biguanide Shock
Borate
Bromine (DPD)
Chlorine, Free (DPD)
Chlorine, Combined
Chlorine, Total (DPD)
Test results that are out of range of the reagent system wil be RED.
accurate. Go SETTINGS>Other Settings to turn the Ranges Enabled feature OFF.
Copper concentrations over 4 ppm will result in low total hardness results.
to exit to the Settings menu.
to move to the next value. After the last value has been chosen
to return to the Settings menu.
Disk
Range
Abbreviation
0–250 ppm
ALK
0–70 ppm
BIG
0–250 ppm
BSK
0–80 ppm
BORATE
0.00–33.00 ppm
BR
0.00–15.00 ppm
FCL
0,00–15.00 ppm
CCL
0.00–15.00 ppm
TCL
to return to the test screen at any time.
to return to the Settings menu.
to exit to the Settings menu.
to exit to the Settings menu.
to exit to the Settings menu.
to return to the Settings menu.
Test Factor
Copper
Cyanuric Acid
Hardness,
Calcium
Iron
pH
Phosphate
Salt
8
and
or
to exit to the Settings menu.
Disk
Range
Abbreviation
0.0–3.0 ppm
COPPER
5–150 ppm
CYA
0–800 ppm
HARD
0.0–3.0 ppm
IRON
6.4–8.6
PH
0–2000 ppb
PHOS
0–5000 ppm
SALT
RED
test results may not be
to
to