Immunity test
IEC 60601
test level
Power frequency
3A/m
(50Hz) magnetic
field
IEC 61000-4-8
Guidance and manufacturer's declaration – electromagnetic immunity for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
The blood pressure monitor (SBM 67) is intended for use in the electromagnetic environment specified below. The customer
or the user of blood pressure monitor (SBM 67) should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF
3 V
IEC 61000-4-6
150 kHz to
80 MHz
Compliance
level
3A/m
Compliance
level
3 V
rms
rms
Electromagnetic environment – guidance
Power frequency magnetic fields should be at levels characte-
ristic of a typical location in a typical commercial or hospital
environment.
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the blood pressure monitor (SBM 67), inclu-
ding cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance.
32