Water Supply Line Connection; Cavitron; System Controls - Dentsply Sirona Cavitron 300 Serie Instrucciones De Uso

Ocultar thumbs Ver también para Cavitron 300 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 34
Main Power Switch
Power Input

7.6 Water Supply Line Connection

• Grasp the Water Supply Line (blue hose) by the end opposite
the quick-disconnect and insert it into the water inlet connector
until fully seated.
• Connect the quick-disconnect to the dental office water supply
or a Cavitron DualSelect Dispensing System.
• Inspect all connections to make certain there are no leaks.
• To remove the water line from the Cavitron System, turn off
the dental office water supply. Disconnect the water supply
line from the dental office water supply. If a quick-disconnect
connector is attached to the end of the hose, relieve the
water pressure by pressing the tip of the connector into an
appropriate container and allow water to drain. To remove the
hose from the system, push on the outer ring of the system's
water inlet and gently pull out the water line.
Press ring to release
water supply tube
.
7.7 Turning the Foot Pedal On/Off
• The foot pedal is packaged with the power OFF. The foot pedal
must be turned on prior to use. (The foot pedal will not charge
in the OFF state).
• The Power button is used to disconnect the battery from the
circuitry for shipping purposes.
• The Power button must be pressed and held for 4 seconds to
disconnect the battery. Likewise the Power button must be
pressed and held for 4 seconds to reconnect the battery.
• If the Foot Pedal power is OFF, pressing the Sync button or
trying to Sync with the Scaler will not turn the Foot Pedal on.
7.8 Charging the Foot Pedal
Ensure the foot pedal power is ON. With the unit ON, plug the
Auxiliary foot pedal cable into the foot pedal and the USB port on
the front of the unit. Allow to charge for up to 4 hours for a full
battery charge.
The unit will operate with the foot pedal while the battery is
charging.
Battery Icon (Foot Pedal Charge)
100% - 75%
75% - 50%
50% - 25%
25% - 0%
7.9 Synchronizing the Foot Pedal
The Tap-On Wireless Rechargeable Foot Pedal supplied with your
system has been factory synchronized with the base unit. Should
a replacement Foot Pedal be necessary, synchronization will be
required prior to system operation. Perform the following steps to
synchronize the Foot Pedal with the base unit.
1.
Turn the Main Power switch located at the rear of the system
to the ON (I) position. The Main screen will appear. This is also
referred to as the Scale screen.
2. Press and hold the Settings icon until the Settings screen
appears.
3. Maintain a distance of no more than 10 feet (3 meters)
between the base unit and Tap-On Foot Pedal during the
synchronization process.
4. Remove any inserts from the handpiece.
5. Ensure the foot pedal power is ON.
6. Press the Sync icon on the Settings screen.
7. Press the red sync button on the bottom of the foot pedal -
hold for at least three seconds.
8. Synchronizing is complete when the Sync icon has stopped
rotating and sound indication occurs.To verify proper
communication, press the Scale icon to return to the Main
screen (scale screen). Depress the foot pedal to the second
position and ensure that Boost is activated (seen in Power
Level icon).

8. Cavitron

300 Series Ultrasonic Scaler
®

8.1 System Controls

Diagnostic Display
See section 8.2 for more details.
Purge
Purge mode is used to flush
water lines. Runs automatically
for 2 minutes (Page 9).
Rinse
Rinse mode is for use
during an ultrasonic
scaling procedure when
lavage is desired with no
cavitation (Page 9).
Ultrasonic Power
Level Control
Slide to select the
desired ultrasonic
power level for operation.
Sliding up increases the
power level and sliding
down decreases the
power level.
Settings Icon
Allows you to adjust
settings for the foot
pedal and screen
brightness. Press and
hold until the Settings
screen appears.
Power On Indicator
English - 7
Tap-On™ Wireless Rechargeable Foot Pedal
Eliminates the need to hold down or pump foot pedal.
See section 8.5 for more details.
Main Power ON/OFF Switch
ON/OFF Switch located
located on the rear of the
system.
Steri-Mate
®
360 Handpiece
Accepts Cavitron
30K
®
Ultrasonic inserts. See
section 8.3 for more details.
Preset Power Modes
Handpiece Holder
Securely holds the system's
handpiece, or cable
connector when handpiece
is not installed.
Lock / Unlock
Press and hold to lock or
unlock the screen. Screen will
appear dimmed when locked.
Detachable Featherweight
Handpiece Cable
Cable can easily be detached
from unit as needed for
replacement.
English - 8