Press the START button and the display will show a flashing calibration value
Place the appropriate test weight in the center of the pan.
Wait for a steady indicator and the display will return to weighing mode.
Calibration is complete
Place a known weight on the pan to check that it is accurate.
If the displayed weight does not match the value of the calibration weight, repeat
the above steps.
3.3.2.9. Cooperation with a computer
Connect the scale to the computer with the RS232 cable: