Attention!
Insert the piston at a slight angle to avoid damage to the lip
seal.
6. Mount the lifting spring (6) and carefully reinsert the piston (5).
7. Replace the outer sleeve (2) and press it down until it audibly
locks in place.
8. Push down the securing ring (3).

9. Cleaning/replacing the valves

9.1. Cleaning/replacing the intake valve
Note:
Valve ball (9) may fall out when valve is removed
1. Unscrew the intake valve (10) using the valve tool (17).
2. Clean or replace the valve.
3. Make sure that valve ball is still in place. To reassemble, handtighten
valve (10) and complete using valve tool (17).
30