micro-TRI-effect
4 Calibration and Check
Calibration on white
A calibration on white is necessary every 30 days. If white calibration is expired,
a red cross is displayed. Click on the icon.
1.
Open snap closure and place instrument on white calibration tile.
2.
Click on instrument icon or press button Operate.
3.
Color and effect measurement on white tile is performed.
4.
Same checks on green tile are performed, see below.
5.
Confirm with checkmark in upper right corner.
After successful calibration and check the red cross turns into a green check-
mark.
Color check on green
A color check on green is recommended every 7 days. This interval is not ob-
served by the instrument and not indicated on the display. Click on the icon.
1.
Open snap closure and place instrument on green color check tile.
2.
Click on instrument icon or press button Operate.
3.
Color and effect measurement on green tile is performed.
4.
Confirm with checkmark in upper right corner.