3.3 Replacing the Reverse Osmosis Module LC240
Follow the instructions below.
Chorus is fitted with Reverse Osmosis (RO) modules.The RO Modules may need replacing if the permeate water purity or
flow rate is not adequate and does not meet predicted or previous performance.
Colour tubes for illustration purposes only.
1. Process button off.
2. Unscrew RO housing.
5. Disconnect RO tubing.
WARNING!
ALL NEW RO MODULES ARE FILLED WITH A PRESERVATIVE SOLUTION TO PREVENT BACTERIAL
CONTAMINATION DURING STORAGE. THE RO MODULES WILL THEREFORE REQUIRE RINSING PRIOR TO USE.
6. Remove and replace RO module.
7. Power ON the systern and water supply.
8. Hold and press process button for
15 second to enter commissioning mode.
(1 hours rinse cycle)
Chorus 3 Operator Manual MANU39996 Version 02 3/17 Page 5
3. Turn off power and water supply.
4. Remove RO housing.
Product outlet
Drain
Inlet