9.
Use of camera
The camera is driven by the software PROVIEW.
You have to run the setup.exe file for the software installation.
At the end of the installation it is possible to start the software.
NOTE: no driver installation is required for the cameras. The software setup procedure automatically
installs all the needed drivers for the correct operation of the camera.
The software's User Manual is available in PDF format within the application itself and can be opened using the
"F1" function key.
You must have Acrobat Reader installed to view the manual.
The manual contains all the operating instructions for using the camera and for the various functions of the soft-
ware.
9.1
WiFi camera connection
The camera is powered via USB cable.
The camera connects to PC only and exclusively with a WiFi connection.
There is no possibility to connect the camera to PC with a USB cable.
1. Connect the USB power cable to the camera.
2. In the Settings of your PC select the WiFi device C-WFR-xxxxx.
The password for connecting the camera is "12345678".
3. Launch the PROVIEW.
4. Manage the camera using the software to control all function.
9.2
Use of the camera
1. Connect the USB power cable to the camera.
2. From "Camera List" panel in the software, click on the camera name.
3. Live image appears.
4. Manage the camera using the software to control all function.
9.3
Parfocality adjustment
In order to have the same focus when observing the spec-
imen through the eyepieces and on the screen/monitor,
please check that the microscope is properly installed and
follow the instructions below.
9.3.1 Biological microscope
1. Use a low power objective and focus the specimen.
2. Switch to the highest dry objective available on the mi-
croscope (40x or 60x) and focus the specimen again.
3. Turn on the live-view on the camera, without changing
the focus on the microscope.
4. While observing the image on the screen/monitor, adjust
the focus by turning the knurled knob on the C-mount
adapter. (Fig. 11)
9.3.2 Stereomicroscope with "C" mount adapter
1. Use a low power magnification and focus the specimen.
2. Reach the highest magnification available using the
zoom knob and then focus the specimen again.
3. Turn on the live-view on the camera, without changing
the focus on the microscope.
4. While observing the image on the screen/monitor, adjust
the focus by turning the knurled knob on the C-mount
adapter. (Fig. 11)