Technical Specifications
Dimensions
Weight
System operating temperature
Docking station storage
and transport humidity
and temperature
Docking station operating
humidity
Electrical Rating
AC adapter
Only use adapters recommended
by HemoCue.
Pollution degree
Over voltage category
Atmospheric pressure
Equipment not suitable for use in the presence of flammable mixtures
The HemoCue 201 DM Docking Station is tested according to IEC 61010-1, IEC 61326-1 and IEC 61326-2-6
Electromagnetic Compatibility
The HemoCue 201 DM Docking Stations complies with emission and basic immunity requirements
described in IEC 61326-1.
The HemoCue 201 DM system complies with the emission and immunity requirements described in
IEC 61326-2-6. When using the analyzer standalone or together with one Primary Docking Station it
has been designed and tested to CISPR 11 Class B. When using the analyzer with a standalone Second-
ary Docking Station or one or several Secondary Docking Stations connected to a Primary Docking
Station it has been designed and tested to CISPR 11 Class A. In a domestic environment it may cause
radio interference, in which case, you may need to take measures to mitigate the interference.
The electromagnetic environment in which the HemoCue 201 DM systems will be operated should be
evaluated prior to operation of the device. Do not use the HemoCue 201 DM systems in close proxim-
ity to sources of strong electromagnetic radiation (e.g. mobile phones and other wireless communica-
tion devices) as these can interfere with the proper operation.
Note: It is the manufacturer's responsibility to provide equipment electromagnetic compatibility informa-
tion to the customer or user.
Note: It is the user's responsibility to ensure that a compatible electromagnetic environment for the
equipment can be maintained in order that the device will perform as intended.
Warranty
The HemoCue 201 DM Docking Station carries a 24-month warranty from the day of receipt.
After the warranty period, service/repair is carried out at fixed prices. Any other use of the System
than recommended by the manufacturer, will void the warranty.
Service and Disposal
The 201 DM Docking Station should be cleaned as recommended under Maintenance in the
Instructions for Use for HemoCue Hb 201 DM Analyzer, HemoCue Glucose 201 DM Analyzer or
HemoCue Glucose 201 DM RT Analyzer, respectively, prior to service or disposal. Consult your local
environmental authorities for proper disposal.
10
206 × 135 × 61 mm (8.10 × 5.30 × 2.40 inches)
566 g (1.24 pounds)
See relevant Instructions for Use
< 90% RH (non-condensing) 0 - 50 °C (32 - 122 °F)
< 90% RH (non-condensing) 15 - 25 °C (59 - 78 °F)
< 75% RH (non-condensing) 26 - 30 °C (79 - 86 °F)
Primary Docking Station: 6 V, 2.5 A
Secondary Docking Station: 6 V, 0.5 A
FW7556M/06: Input: 100V~ – 240 V AC/50–60 Hz/400-200 mA
Output: 6 V DC/2.5 A
FW8001/06: Input: 100V~ – 240 V AC/50–60 Hz/400-200 mA
Output: 5.9 V DC/3.0 A
2
II
700 hPa to 1060 hPa