c) Gently press the Secondary Docking Station (5) downwards onto the
Primary Docking Station (6), fitting the Female Connector and the Male
connector together.
Note! The Docking Stations must be securely connected to each other
in order to ensure good electrical contact.
d) If more Secondary Docking Stations are to be added, remove the
Side Trim from the previously added Docking Station and the bottom
trim of the new Secondary Docking Station, follow step a) and b). Re-
peat the procedure in step c) with the new Secondary Docking Station.
5
6
Important! If not connected to another Docking Station the Side Trim
(2) on the Docking Station should always be used to protect the
.
Connector (1)
Docking Station LED
Steady green light
A steady green light from the LED (1) indicates that the Docking Station
(3) is receiving power and that the Analyzer Battery is fully charged.
Flashing green light
A flashing green light from the LED (1) indicates that the Analyzer
Battery in the docked Analyzer (2) is charging.
Steady red light
A steady red light from the LED (4) indicates an internal error within the
Docking Station. See section Troubleshooting.
Note! Not valid for unlabelled Secondary Docking Stations (non-SDS
Flashing red light
A flashing red light from the LED (4) indicates an external communica-
tion error. See section Troubleshooting.
Note! Not valid for unlabelled Secondary Docking Stations, (non-SDS
).
+
+
).