Essential Performance
When operated at maximum suction (Level 8, Mode 3) the vacuum pressure must not
exceed its specification.
EN 60601-1-2 STANDARD TEST REQUIREMENTS
Immunity Test Levels for ENCLOSURE PORT
PHENOMENON
Electrostatic Discharge
Radiated RF EM fields
Proximity fields
from RF wireless
communications
equipment
RATED power
frequency magnetic
fields
d) e)
Immunity Test Levels for Input AC Power PORT
PHENOMENON
Electrical fast
transients/bursts
a) l) o)
Surges
a) b) j) o)
Line-to-line
Surges
a) b) j) k) o)
Line-to-ground
Conducted
disturbances
induced by RF fields
c) d) o)
Voltage dips
f) p) r)
Voltage Interruptions
f) i) o) r)
Conducted Emission Test Level
FREQUENCY RANGE
(MHz)
0,15 – 0,5
0,5– 5
5 – 30
BASIC EMC
STANDARD OR
TEST METHOD
IEC 61000-4-2
IEC 61000-4-3
a)
IEC 61000-4-3
IEC 61000-4-8
BASIC EMC
STANDARD OR
TEST METHOD
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
IMMUNITY TEST LEVELS
Professional Healthcare
Facility Environment
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
3 V/m
f)
80 MHz – 2,7 GHz
80 % AM at 1 kHz
IMMUNITY TEST LEVELS
Professional Healthcare
Facility Environment
100 kHz repetition frequency
± 0,5 kV, ± 1 kV, ± 2 kV
3 V
m)
0, 15 MHz – 80 MHz
6 V
in ISM bands between
m)
0,15 MHz and 80 MHz
80 % AM at 1 kHz
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
70 % UT ; 25/30 cycles
0 % UT ; 250/300 cycles
QUASI-PEAK
(dBµV)
56.91 - 37.86
37.86 - 36
36 - 40
Home Healthcare
Environment
± 8 kV contact
80 MHz – 2,7 GHz
b)
80 % AM at 1 kHz
c)
See 8.10
30 A/m
g)
50 Hz or 60 Hz
Home Healthcare
Environment
± 2 kV
± 0,5 kV, ± 1 kV
0, 15 MHz – 80 MHz
6 V
in ISM and amatuer
m)
radio bands between 0,15
n)
MHz and 80 MHz
e)
80 % AM at 1 kHz
0 % UT ; 0,5 cycle
g)
0 % UT ; 1 cycle and
h)
Single phase: at 0°
h)
AVERAGE
(dBµV)
38 - 25
25 - 26
26 - 27
10 V/m
f)
b)
c)
3 V
m)
n)
e)
q)
19