Signum: Installing in the Scale
Installation of the interface module in the scale is only required if the customer did not
order the respective data output port to be fitted in the factory.
Notes
§ The interface module should only be installed by a trained and authorized Minebea Intec
technician.
§ IP65 protection:
Use the appropriate connecting cable with a cable gland designed for the interface
module in question (see "Accessories").
Disconnect the equipment from the electricity supply prior to commencing work.
!
Improper installation invalidates the applicable warranty.
!
!
Perform all installation work that affects the IP65 protection rating with extreme care.
§ Remove the rear wall panel: Remove the 6 screws from the rear wall panel.
§ Carefully remove the rear wall panel. (The RS-232 interface is connected to the digital
PCB by a ribbon cable.)
§ Remove the RS-232 interface (including the accompanying PCB) from the rear wall panel:
Remove the 2 screws from the connector mounting.