Contact Protection Measures - WIKA switchGauge 821 Manual De Instrucciones

Ocultar thumbs Ver también para switchGauge 821:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5. Commissioning, operation
Seal the cable entry with the appropriate approved cable glands.
Cable gland design
Standard
Install the connection cables securely.

5.3 Contact protection measures

Mechanical contacts must not exceed the specified electrical values for switching
current, switching voltage and switching power independent of each other, not
even for a short time only.
For capacitive or inductive loads we recommend one of the following protective
circuits:
Inductive load with DC voltage
With DC voltage the contact protection can be achieved via a free-wheeling di-
ode, connected in parallel to the load. The polarity of the diode must be arranged
so that it closes when the operating voltage is on.
Example:
Contact protec-
tion measure with
free-wheeling
diode
Operating instructions pressure gauges with switch contact models 821, 851 or 830 E
Only use
cable with a
diameter of
7 ... 13 mm
Contact
Instruments > 4 kg
Diode
Load
Only use
cable with a
diameter of
5 ... 10 mm
15
EN

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Switchgauge 851Switchgauge 830 e

Tabla de contenido