Important information
!
Do not use the defibrillator to shock the PACER or ECG terminals.
!
When connecting to the DEFIB terminals, 4mm shrouded non-stackable plugs must
be used.
!
The DEFIB terminals (red and black) are rated to 5kV/100A/600J MAX.
!
The PACER terminals (yellow) are rated to 275V MAX.
!
The UniPulse 400 is designed for use by adequately trained technical personnel only.
!
No unauthorised user modifications of the instrument.
!
Do not use the instrument in an application beyond the published specifications.
!
Use extreme caution when working with voltages greater than 30 Volts, for example
a DEFIB shock.
!
No user serviceable parts.
!
Do not use the Uni-Pulse 400 with the 18V PSU attached unless it is connected
directly to mains power.
Where safe operation of the Uni-Pulse is no longer possible it should be immediately
!
shut down and secured to prevent accidental operation. It must be assumed that safe
operation is no longer possible:
-
if the instrument or leads show any sign of damage or
-
the instrument does not function or
-
after long periods of storage under adverse environmental conditions or
-
do not use the UniPulse 400 in wet conditions or surrounded in explosive
gases or dust.
!
The UniPulse 400 is designed to analyse Defibrillators, Automated External
Defibrillators (AED's) and Pacers, and allows Electrocardiogram (ECG) simulation.