Sensor offset
Although the NeuLog pH sensor does not need to be
calibrated prior to each use, it is recommended to offset
its probe.
Materials needed:
pH 7 Buffer Solution (for qualitative measurements,
distilled or even tap water can be used instead)
Procedure:
1. Connect the pH sensor to a voltage source (USB-200
or BAT-200).
2. Remove the pH sensor probe's tip from the plastic
storage solution bottle.
3. Rinse the pH sensor's tip with distilled water.
4. Insert the pH sensor's tip into a pH 7 buffer; if no
buffer is available, you can use tap water instead
(although it is not as accurate).
5. Press and hold the pH sensor's push button for 3
seconds.
V2018.01
19