Maintenance and storage
Never submerge the NeuLog sensor's plastic body in
any liquid.
Never submerge the NeuLog plastic body in any
liquid.
Do not allow liquid into the pH sensor's body.
After use, gently wipe away any foreign material from
the pH sensor plastic body.
Rinse the electrodes with distilled water before and
after measuring a sample. Never wipe the electrode to
remove excess water – wiping can create static
charges that interfere with correct pH measurement.
When storing your sensor, always keep your pH
electrode wet. We recommend that you store your
electrode in either a commercially prepared storage
solution, or a 1:1 solution of pH 4 buffer and 4M KCl.
If you do not have the pH 4 buffer + 4M KCl solution
available, a pH 4 buffer can be used. If you do not
have a pH 4 buffer available, a pH 7 buffer can be
used.
Do not store the electrode in distilled or deionized
water – this will cause ions to leach out of the glass
bulb and render your electrode useless.
Store in a box at room temperature out of direct
sunlight.