and acoustic effects of the irrigating fluid, in most cases, is
adequate to remove even the most tenacious calculus.
• The Blue Zone
, an extended low power range, is marked on
the dial in blue.
• Set the System's power adjustment knob to the lowest power
setting for the application and the selected insert.
• Only adjust the system's power adjustments knob with the
insert outside the patients mouth.
9.3 Patient Comfort Considerations
Reasons for sensitivity
• Incorrect tip placement. Point should be directed away from
root surfaces.
• Not keeping the tip in motion on the tooth. Do not allow the
insert to remain in a static position on any one area of the tooth.
Change the insert's path of motion.
• Applying pressure. Use extremely light grasp and pressure,
especially on exposed cementum.
• If sensitivity persists, decrease power setting and/or move
from the sensitive tooth to another and then return.
Section 10: System Care
10.1 Daily Protocol
Start-up procedures at the beginning of the day
1.
Switch ON the dental treatment unit (refer to instruction manual
of treatment unit).
®
2.
Remove the Cavitron
cable connector assembly from
handpiece holder and set power adjustment to top of Blue Zone.
3.
Connect a sterilized Steri-Mate
connector assembly. (Refer to the Infection Control information
booklet for sterilizing instructions.)
4.
Adjust the lavage control to maximum.
5.
Hold the handpiece (without an insert installed) over a sink
or drain. Activate the foot control and flush water through
handpiece for at least two minutes.
6.
Place a sterilized insert into the handpiece using a gentle push-
twist motion. (Refer to the Cavitron Ultrasonic Inserts Directions
for Use for cleaning and sterilizing instructions.)
7.
Activate the ultrasonics and adjust the power and lavage control
to your preferred operating positions.
Between patients
1.
Remove any ultrasonic inserts and the Steri-Mate
Clean and sterilize the handpiece and all inserts used during the
procedure. (Refer to the Infection Control Information booklet and
the Cavitron Ultrasonic Inserts Directions for Use for cleaning and
sterilizing instructions.)
2.
Clean and disinfect the handpiece cable assembly by applying
a water-based disinfectant solution*, carefully following the
instructions provided by the solution manufacturer. To clean
®
handpiece to the cable
®
handpiece.
ENGLISH • 13