Switching on the scale
Start weighing
CAUTION!
Patients can be injured if their limbs get pinched
The patient's hands and arms can be pinched
between the seat and the lower frame or
between the seat and the armrests.
– Make sure that the patient's hands and arms
are on the armrests or in his lap during the
weighing process.
ATTENTION!
Incorrect measurement due to force shunts
If accessories (e.g. the power pack pouch) or
articles of clothing get caught between the seat
and the frame, the weight will not be measured
correctly.
If the patient's feet are not on the footrests, the
weight will not be measured correctly.
– Make sure that there are no accessories or
articles of clothing between the seat and
frame.
– Make sure that the armrests are folded down
during the weighing process.
– Make sure that the patient's feet are on the
footrests during the weighing process.
Press the Start key.
The display will read
through all elements of the display.
The scales are ready for operation when the display
0. 0 0
reads
.
If the device is being operated with a power supply
unit, the symbol
1. Set the wheel brakes of the rear transport rollers.
se[A
and then quickly run
will appear in the display.
Operation • 67
English