Command List
This section lists all commands and queries implemented in this instrument. The commands are
listed in alphabetical order within the function groups.
The following nomenclature is used:
<rmt>
<nr1>
<nr2>
Set-up Commands
V <nr2>
I <nr2>
ON
OFF
Readback Commands
V?
I?
VO?
IO?
OUT?
M?
10
<RESPONSE MESSAGE TERMINATOR>
A number with no fractional part, i.e. an integer.
A number in a fixed point format, e.g. 11.52, 3.61, etc.
Set the voltage to <nr2>. The value of <nr2> must be in Volts; no multipliers are
allowed. If the value of <nr2>, after rounding, is outside the range of the output an
error will be generated.
Set current limit to <nr2>. The value of <nr2> must be in Amps; no multipliers are
allowed. If the value of <nr2>, after rounding, is outside the range of the output an
error will be generated.
Sets the DC output on.
Sets the DC output off.
Returns the set output voltage in Volts in <nr2> numeric format.
The syntax of the response is
Example:
If the set output voltage is 12.55 Volts the response to
the command V? will be V 12.55<rmt>.
Returns the output current limit in Amps in <nr2> numeric format. The syntax of the
response is
Example:
If the current limit is 1.00 Amps the response to the
command I? will be I 1.00<rmt>.
Reads and returns the actual output voltage in Volts in <nr2> numeric format. The
syntax of the response is
Example:
If the actual output voltage is 12.55 Volts the
response to the command VO? will be V12.55<rmt>.
Reads and returns the actual output current in Amps in the <nr2> numeric format.
The syntax of the response is
Example:
If the output current is 0.93 Amps the response to
the command IO? will be A0.93<rmt>.
Returns status of the output as
or
Returns output mode status as
or
, CR followed by LF.
V<nr2><rmt>
I<nr2><rmt>
V<nr2><rmt>
I<nr2><rmt>
OUT ON<rmt>
OUT OFF<rmt>
M CV<rmt>
M CC<rmt>