KPS-LX10 • Digital light meter
3-3 Correction Factor
• Mercury Lamp .................................... x1.1
• Fluorescent Lamp .............................. x1.0
• Incandescent Light............................. x1.0
• Daylight .............................................. x1.0
4.
SPECTRUM FOR PHOTO DETECTOR
5.
MEASURING PROCEDURE
• Power-up: Press the POWER key to turn the meter on.
• Selecting the LUX scale: Set the Function switch to desired LUX range.
• Face the photo detector to light source in a horizontal position.
• Read the illuminance value from the LCD display.
• Data-Hold mode: Press the HOLD key to select Hold mode. When
• HOLD mode is selected, the illuminance meter stops all further measure-
ments. Press the HOLD key again to cancel HOLD mode.
• Then it resumes normal operation.
• When the measurement is completed, turn the power off.
Note:
If display indicates one or more leading zeros, user has toshift the Function
switch to the next lower range scale to improve resolution and accuracy.
6.
MAINTENANCE
6-1 General maintenance
• To keep the instrument clean, wipe the case with a dry cloth and detergent,
do not use abrasives or solvents.
• The white plastic disc on the top of the detector should be cleaned with a
damp cloth when necessary.