If no signal is detected, the LCD will show EF, NCV indicator light doesn't flashes and the buzzer
is no sound.
According to the detected signal strength, LCD displays different horizontal lines.
When the signal is strongest, LCD displays four horizontal lines, when the signal is weakest,
only one line.
Meanwhile, NCV indicator light flashes, buzzer make different sound.
Note: The sensing level varies with the distance between the sensing part and the measured AC
power cord.
Note: The detector is designed with high sensitivity. Static electricity or other sources of energy may
randomly trip the sensor. This is normal operation.
The absence of a voltage indication in the NCV function does not mean that there is no voltage.
To confirm the absence of a voltage, use a VAT.
The presence of other voltages nearby can trigger non-contact voltage detection.
5.1. SPECIFICATIONS
Note: Intrinsic uncertainty is stated at 18 to 28°C (65 to 83°F) and less than 75%RH.
Note: Intrinsic uncertainty is expressed as a percentage of reading (R) plus a number of digits (ct):
±(a% R + b)
Function
AC True RMS Voltage
All AC voltage ranges are specified from 5% of range to 100% of range.
AC Voltage Bandwidth: 50 to 60 Hz (All wave); 50 Hz to 1 kHz (Sine wave).
DC Voltage
5. SPECIFICATIONS
Range
4.000 V
40.00 V
400.0 V
600 V
400.0 mV
4.000 V
40.00 V
400.0 V
600 V
19
Resolution
1 mV
10 mV
100 mV
1 V
0.1 mV
1 mV
10 mV
100 mV
1 V
Intrinsic uncertainty
±(1.2% + 3 ct)
±(1.2% + 5 ct)
±(0.5% + 8 ct)