5. Commissioning, operation
5.4.2.2
Writing new parameters
Take note of the new communication parameters before writing them, as the parameters will be required again for any
new access to the gas density sensor.
Write the new values in the right fields (below the button "Write on instrument").
Designation
Valid values
Tag number
16 characters in ASCII code
Address
1 ... 247
Baud rate
1,200 ... 115,200
Parity
None, even
By pressing the button "Write on instrument" the data in the fields is transmitted to the instrument register. To finish the
writing operation, interrupt the voltage supply of the gas density sensor after the transmission before restoring it.
Afterwards, during the reading operation, the entered data becomes visible on the left-hand side.
If Windows
must be changed to English (USA), since otherwise, communication problems might occur.
5.4.2.3
Data logger
The data logger is used for recording measured values over a certain time span.
WIKA operating instructions gas density sensor, model GD-20
All manuals and user guides at all-guides.com
is used with non-Latin character sets (e.g. Chinese), the area settings of the system control
®
EN
55